ENTRY SKIN DOSE REDUCTION IN AN INLINE MRI-LINAC USING AN ELECTRON CONTAMINATION DEFLECTOR COUPLED WITH A HELIUM VOLUME

Authors

  • Olalekan Adekunle Adeyemo Physics, Dunster Business School Author

DOI:

https://doi.org/10.61841/kt28ca90

Keywords:

MRI-Linac, electron contamination, entry-surface dose, electron contamination deflector, helium volume, radiation transport, computational dosimetry, Monte Carlo simulation.

Abstract

While magnetic resonance imaging (MRI) and linear accelerators (MRI-Linac) are used together for better visualizing soft-tissue prior to radiotherapy, the dosimetric issues are that secondary electron transport and entry-surface dose distribution are challenging. In MRI-Linac systems with inline beam path, dose to the skin may be overestimated due to the presence of the electron contamination in the beam. Four configurations are studied: a baseline configuration, an “electron contamination deflector (ECD)”, a configuration with helium-filled volume and a combined ECD–helium. A computational radiation transport framework written in python was used to compare electron-fluence outputs, electron depth dose distributions, beam-quality parameters and dose-validation metrics. It took 200,000 primary photons in the combined configuration, and 50,000 in the baseline (ECD-only) and helium-only configurations. The numbers of surface electrons counted were: 34, 22, 16 and 27. The electron-arrival rates normalized to the number of primary photons were 6.80 $\times$ 10$^ {-4} $, 4.40 $\times$ 10$^ {-4} $, 3.20 $\times$ 10$^ {-4} $ and 1.35 $\times$ 10$^ {-4} $ electrons per primary photon. The normalized electron-arrival rate of this combined configuration was the smallest, and 80.1% descriptive reduction was obtained compared to the baseline. However, the dose-validation results of the combined configuration showed a Pearson correlation of −0.072 with the reference dose distribution and 0% gamma pass rate under the 3%/3 mm criteria. The results represent differences in the reported rates of arrival of electrons at the simulated set-ups. The results from the dose validation indicate that the dose calculated, and the reference distributions agree only to a certain degree. A comparison is made of the ECD helium configuration in the computational framework studied and their dosimetric outputs and electron-fluence and the validity of their results are interpreted.

References

Lagendijk, J. J., Raaymakers, B. W., Van den Berg, C. A., Moerland, M. A., Philippens, M. E., & Van Vulpen, M. (2014). MR guidance in radiotherapy. Physics in Medicine & Biology, 59(21), R349-R369.

Raaymakers, B. W., Jürgenliemk-Schulz, I. M., Bol, G. H., Glitzner, M., Kotte, A. N. T. J., Van Asselen, B., ... & Lagendijk, J. J. W. (2017). First patients treated with a 1.5 T MRI-Linac: clinical proof of concept of a high-precision, high-field MRI guided radiotherapy treatment. Physics in Medicine & Biology, 62(23), L41-L50.

Pojtinger, S., Kapsch, R. P., Dohm, O. S., & Thorwarth, D. (2019). A finite element method for the determination of the relative response of ionization chambers in MR-linacs: simulation and experimental validation up to 1.5 T. Physics in Medicine & Biology, 64(13), 135011.

Begg, J., Jelen, U., Moutrie, Z., Oliver, C., Holloway, L., Brown, R., & ACPSEM Magnetic Resonance Imaging Linac Working Group (MRILWG). (2023). ACPSEM position paper: dosimetry for magnetic resonance imaging linear accelerators. Physical and Engineering Sciences in Medicine, 46(1), 1-17.

Hackett, S. L., van Asselen, B., Wolthaus, J. W., Bluemink, J. J., Ishakoglu, K., Kok, J., ... & Raaymakers, B. W. (2018). Spiraling contaminant electrons increase doses to surfaces outside the photon beam of an MRI-linac with a perpendicular magnetic field. Physics in Medicine & Biology, 63(9), 095001.

Keyvanloo, A., Burke, B., St. Aubin, J., Baillie, D., Wachowicz, K., Warkentin, B., ... & Fallone, B. G. (2016). Minimal skin does increase in longitudinal rotating biplanar linac-MR systems: examination of radiation energy and flattening filter design. Physics in Medicine & Biology, 61(9), 3527-3539.

Downloads

Published

2026-11-19

How to Cite

Adeyemo , O. A. (2026). ENTRY SKIN DOSE REDUCTION IN AN INLINE MRI-LINAC USING AN ELECTRON CONTAMINATION DEFLECTOR COUPLED WITH A HELIUM VOLUME. Journal of Advanced Research in Medical and Health Science, 12(6), 8-12. https://doi.org/10.61841/kt28ca90

Similar Articles

11-20 of 55

You may also start an advanced similarity search for this article.